High-low temperature laboratory, constant temperature and humidity room, walk-in laboratory design
Â
High-low temperature laboratory, constant temperature and humidity room, walk-in laboratory design
product name
Walk-in constant temperature and humidity test chamber
Product number
HG-BW-018
Sample limit
This test equipment prohibits the test of flammable, explosive, volatile substance samples and the test or storage of corrosive substance samples, biological samples, and strong electromagnetic emission source samples.
Volume, size and weight
Content
custom made  Â
Test room size
4300MM × 4650MM × 2200MM (Deep × width × height) can be customized according to demand
Performance
Functional requirements
1, (air) indoor space cooling, heating, humidification, dehumidification; according to temperature, humidity settings, indoor space to set the temperature, humidity;
2, to provide independent cooling water device;
3. Each function must work independently.
Test environmental conditions
Ambient temperature +25°C, relative humidity ≤85%, no test sample in the test chamber
Test Methods
GB / T 5170.5-1996 electric and electronic products environmental testing equipment basic parameters of test equipment damp heat test method
â–²Temperature range
20°C— +25°C
â–²Humidity range
25 %RH-90%RH
Temperature and humidity curve
â–²Temperature uniformity
When there is blowing, it is ≤± 2°C. (In the working area test, the work area is at a distance of 300 mm from the four sides and the top and bottom walls, and the temperature uniformity is not guaranteed when there is no blowing in the box.)
â–² humidity uniformity
≤± 2.5% when there is a blast (in the working area, the working area is the area of ​​300 mm from the four sides and the top and bottom walls of the work area. If there is no blast, the humidity uniformity is not guaranteed and the humidity cannot be measured.)
â–² warm up time
+0°C~+50°C ≤65min
Note: The heating time is the performance of no-load operation in high-temperature room
â–²Cooling time
+20°C → -40°C≤60min
Non-linear
Satisfy the test standard
1. GB/T10589-1989 cryogenic test chamber technical conditions;
2.GB/T10586-1989 hot and humid test chamber technical conditions;
3.GB/T10592-1989 high and low temperature test box technical conditions;
4. GB2423.1-89 low temperature test Aa, Ab;
5. GB2423.3-93 (IEC68-2-3) constant heat test Ca;
6.MIL-STD810D method 502.2 ;
7.GB/T2423.4-93 (MIL-STD810) method 507.2 procedure 3 ;
8.GJB150.9-8 Damp heat test;
9. GB2423.34-86, MIL-STD883C method 1004.2 temperature and humidity combined cycle test.
Structure
Thermal insulation structure
External materials: High-quality carbon steel plates with electrostatic color spray treatment on the surface. (Cold-rolled steel plate t0.5mm (standard color) (cold-rolled steel plate t0.5+ polyamine standard insulation board)
Internal Materials: Wall Plate , Top Plate Stainless Steel Plate (SUS304)
Base plate Stainless steel plate (SUS304)
Observation window: A 300×450cm electric-thermal-membrane moisture-proof temperature-observation observation window (with glass cover) is opened on the door of the box.
Insulation material: high temperature resistant hard polyurethane foam (T100mm).
Cable hole: box cable hole, diameter Ф50 and equipped with a sealed soft plug.
Air conditioning channel
Fans, heaters, evaporators, dampers, temperature sensors
Test chamber door
Single open
Circulation fan
Centrifugal
control panel
Controller display , fault indicator, timer, over-temperature protection setter, total power switch
Machinery room
Refrigeration unit, connection and drainage device, exhaust fan
Distribution control cabinet
Main power leakage circuit breaker, controller, distribution board, cooling fan, high and low temperature circulation fan motor
Heater
Nickel-chromium alloy heating wire heater
Heater control mode: no contact and other periodic pulse width adjustment, SSR (solid state relay)
Power line hole and drain hole
Located on the back of the test chamber body
Laboratory appearance
(See picture)
Cooling System
Way of working
Adopts the leading domestic mechanical compression binary refrigeration technology
â–²Refrigeration compressor
Imported compressor
Evaporator
Finned tube heat exchanger
Condenser
Coil heat exchanger
Throttling device
Thermal expansion valve, capillary
Evaporative condenser
Stainless steel brazed plate heat exchanger
Refrigerator control method
The control system's PLC (Programmable Logic Controller) automatically adjusts the operating conditions of the chiller based on the test conditions
Compressor return air cooling circuit
Energy regulation circuit
Condensate pressure regulator
Refrigerant
Environmental protection refrigerants with zero ozone depletion index
Cooling System
Compressor cooling, cooling system to increase the plate for cooling, plate replacement provided by the purchaser; hot side of the laboratory cooling system electromagnetic cutting, controlled by the switch on the control panel off, cold side users have other uses, and look forward to communicating with the user How much heat.
other
Compressor cooling fan
Cooling water Filter
Electric control system
Controller model
LCD touch programmable color controller
Features
Independent research and development of dedicated controllers for cooling inverter technology.
(In view of considering that the purchaser can not provide other special experimental conditions in detail due to different experimental needs, the equipment has secondary development functions; it can provide PLC control technical support and development services.)
monitor
Color display
Operation mode
Program mode
Setting method
Chinese menu, touch screen mode input
Program capacity
100 programs. Maximum 1000 steps per program; 9999 cycles per program
Predetermined area
Temperature limit +200°C; Temperature limit -99°C (controller)
display resolution
Temperature: 0.1°C
Time: 0.1min
enter
PT100
Communication function
RS-232C
control method
Anti-integral saturation PID, fuzzy algorithm
BTHC Balanced Temperature and Humidity Control
Curve recording function
Save device sampling values ​​and sampling time
Affiliated functions
Fault alarm and cause, processing prompt function, fault record, over-temperature protection, upper and lower limit temperature protection, power failure protection, sensor up/down wind selection, test pause, alarm output, time signal output, test end output, temperature reach output, timing start And automatic stop function, self-diagnosis function
Software use environment
(user provided)
IBM PC compatible, PII and above CPU, 128M or more memory, Simplified Chinese Windows2000 or Simplified Chinese Windows XP operating system
Temperature and humidity measurement
PT100 platinum resistance
Safety protection device
Cooling System
Compressor overpressure
Overheated compressor
Compressor overcurrent
Circulating cooling water supply under pressure
Laboratory
Laboratory temperature is too high, too low protection (controller built-in)
Laboratory temperature too high protection
Laboratory door protection
other
Total power phase sequence and phase loss protection relay
Earth leakage circuit breaker, sample power supply control terminal, external alarm output terminal
Over-temperature protection switch
Circulation fan thermal relay, motor temperature switch
Compressed air pressure switch, exhaust valve
Other configuration
Power cable
Total power leakage circuit breaker
Five-wire (four-core cable + protective grounding cable) 1 cable (4 meters long)
AC380V, 50Hz
Three-phase four-wire + protection ground
transport
The test room is assembled by library board and split transport (steam transport)
Conditions of Use
The user guarantees the following conditions
Environmental conditions
Temperature: 5°C~35°C
Relative humidity: ≤85%
Circulating cooling water
(water temperature, water pressure)
Recirculating cooling water temperature: ≤ +30 °C
Recirculating cooling water supply pressure: 0.25MPa ~ 0.4MPa   
The design and construction of the cooling water piping system shall ensure that the pressure at the inlet of the refrigerator at the rated flow is 0.25MPa to 0.4MPa , and the pressure drop at the outlet of the refrigerator to the cooling tower is not more than 0.05MPa.
flow
7t/h
Hose fittings and other
One pair of DN40 threaded joints for refrigeration unit
In accordance with GB 50050-1995 industrial circulating cooling water treatment design specification
It is recommended to install an electrostatic water processor (scale removal, algae removal) on the waterway.
Supply conditions and power
AC380V Â Â Three-phase four-wire + protection ground
Voltage allowable fluctuation range: AC (380± 38 )V
Allowable frequency fluctuation range: (50 ± 0.5 ) Hz
The grounding resistance of the protective ground is less than 4Ω; the TN-S mode power supply or the TT mode power supply
Users are required to configure the air or power switch of the corresponding capacity on the installation site. This switch must be used independently for this equipment.
Maximum power
45KW
Requirements for the storage environment
When the equipment is not working, the temperature of the environment should be kept within 0°C~+45 °C
When the ambient temperature is lower than 0°C, the water remaining in the equipment should be drained.
In order to prevent the water inside the pipe from freezing, the pipe may get worse.
    (Note: If there is any deviation from the plus item in the above equipment , the bidder shall indicate and indicate in the bid document, and the corresponding score will be deducted in the score. The reference brand in the demand is for reference only, and the bidder can provide better than the reference brand. Technical parameters of the brand.)